Owner's Manual

9
Intended Use
This system is a 2 in 1 system designed to measure blood glucose
outside the human body (in vitro diagnostic use) and to measure
blood pressure non-invasively. It is intended for use at home and in
clinical settings as an aid to monitoring the effectiveness of diabetes
control. It uses fresh whole blood samples from the nger, palm,
forearm and upper arm.The device is intended for use of all ages for
home and professional use by one with a good understanding of the
operation instruction, where the patient may be an operator.
It shall not be used for the diagnosis of diabetes and hypertension.
This system has voice function, but is not designed for the visually
impaired.
Test Principle
Your system measures the amount of sugar (glucose) in whole blood.
The glucose testing is based on the measurement of electrical current
generated by the reaction of glucose with the reagent of the strip.
Your monitor measures the current, calculates the blood glucose level,
and displays the result. The strength of the current produced by the
reaction depends on the amount of glucose in the blood sample.
The blood pressure is measured non-invasively at the arm based on
the Oscillometric method.
This device is NOT able to take measurements in the presence of
common arrhythmia, such as arterial or ventricular premature beats or
arterial brillation. It may produce reading errors.