Owner's Manual

49
Important Control Solution Information
Use only our control solutions with your monitor.
Do not use the control solution beyond the expiry date or 3 months
after rst opening. Write the opening date on the control solution
vial and discard the remaining solution after 3 months.
It is recommended that the control solution test should be done at
room temperature (20°C to 25°C / 68°F to 77°F). Make sure your
control solution, monitor, and test strips are at this specied
temperature range before testing.
Shake the vial before use, discard the rst drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
Store the control solution tightly closed at temperatures between
2°C and 30°C (36°F and 86°F). Do NOT freeze.