Owner's Manual
22
WARNING
When doing the control solution test, you have to mark it so that the test
result will NOT mix with the blood glucose TEST RESULTS be stored in the
memory. Failure to do so will mix up the blood glucose test results with the
control solution test results in memory.
Step 3
Apply Control Solution
Shake the control solution vial thoroughly before use. Squeeze out a
drop and wipe it off, then squeeze another drop and place it on the tip
of the vial cap.
Hold the monitor to move the absorbent hole of test strip to touch the
drop. Once the conrmation window lls completely, the monitor will
begin counting down.
To avoid contaminating the control solution, do not directly apply
control solution onto a strip.