FORA TD-3261G Blood Glucose Plus Blood Pressure Monitoring System Ownerʼs Manual
English Dear TD-3261G System Owner: Thank you for purchasing the TD-3261G Blood Glucose plus Blood Pressure Monitoring System. This manual provides important information which helps you to operate this system smoothly. Before using this product, please read the following contents thoroughly and carefully. According to the clinical studies from American Diabetes Association, elevated blood pressure often accompanies adult diabetes patients.
IMPORTANT SAFTY INSTRUCTIONS READ BEFORE USE 1. Use this device ONLY for the intended use described in this manual. 2. Do NOT use accessories which are not specified by the manufacturer. 3. Do NOT use the device if it is not working properly or damaged. 4. Do NOT use the equipment where aerosol sprays are being used or where oxygen is being administered. 5. This device does NOT serve as a cure for any symptoms or diseases. The data measured is for reference only.
10. Use of this instrument in a dry environment, especially if synthetic materials are present (synthetic clothing, carpets etc.) may cause damaging static discharges that may cause erroneous results. 11. Do NOT use this instrument in close proximity to sources of strong electromagnetic radiation, as these may interfere with the correct operation. 12. Proper maintenance and periodically control solution test are essential to the longevity of your device.
TABLE OF CONTENTS IMPORTANT SAFETY PRECAUTIONS 04 BEFORE YOU BEGIN 08 Important Information 08 Intended Use 09 Test Principle 09 Monitor Overview 10 Screen Display 12 Test Strip 13 Features 14 SETTING THE MONITOR 15 TESTING YOUR BLOOD GLUCOSE 19 The Four Measuring Modes QUALITY CONTROL TESTING 19 21 Performing a Control Solution Test 21 TESTING WITH BLOOD SAMPLE 24 Preparing the Lancing Device for Blood Testing 24 Preparing the Puncture Site 24 Alternative Site Testing 26 P
TESTING YOUR BLOOD PRESSURE 30 Before Measurement 30 Taking a Single Measurement 33 Average Mode Measurement 34 Auto Cycle Mode 35 MONITOR MEMORY 37 Reviewing Test Results 37 Reviewing Day Average Results 38 DATA TRANSMISSION VIA BLUETOOTH 40 DATA TRANSMISSION VIA 4G 42 MAINTENANCE 45 Battery 45 Using AC Adapter 46 Caring for Your Monitor 47 Caring for Your Test Strips 48 Important Control Solution Information 49 SYSTEM TROUBLESHOOTING 50 Result Readings 50 Error Messag
BEFORE YOU BEGIN Important Information • Severe dehydration and excessive water loss may cause readings which are lower than actual values. If you believe you are suffering from severe dehydration, consult a healthcare professional immediately. • If your blood glucose results are lower or higher than usual, and you do not have symptoms of illness, first repeat the test. If you have symptoms or continue to get results higher or lower than usual, follow the treatment advice of your healthcare professional.
Intended Use This system is a 2 in 1 system designed to measure blood glucose outside the human body (in vitro diagnostic use) and to measure blood pressure non-invasively. It is intended for use at home and in clinical settings as an aid to monitoring the effectiveness of diabetes control. It uses fresh whole blood samples from the finger, palm, forearm and upper arm.
Monitor Overview Front 1 Back 10 9 11 2 3 4 5 6 7 8 Side 13 12 18 Pressure Cuff 14 15 16 10 17
1 TEST SLOT 10 2 3 DISPLAY SCREEN RF BUTTON Assists you with wireless data transmission. 4 5 6 9 13 Enter the monitor memory. 15 ON/OFF BUTTON 16 AVERAGE BUTTON Speaker RF INDICATOR AC/DC ADAPTER PORT Connect to a power supply. 14 Assists you with blood pressure average mode. 8 Eject the used strip by pushing up this button. M BUTTON To start a single blood pressure measurement. 7 BATTERY COMPARTMENT 12 STRIP EJECTOR 11 S BUTTON Enter and confirm the monitor settings.
Screen Display 20 19 1 18 2 17 3 16 4 15 5 14 6 13 7 12 8 11 9 10 1 2 3 4 5 6 7 8 9 10 12 Date Systolic Pressure Symbol Measurement Mode Systolic Pressure Value Diastolic Pressure Symbol Diastolic Pressure Value Volume Symbol Face/Low/High Symbol Ketone Warning Heart Symbol - for IHB Detection (Irregular Heart Beat) 11 12 13 14 15 16 17 18 19 20 Pulse Rate Day Average Result Indicator Low Battery Symbol Memory Symbol Unit for Blood Pressure Unit for Blood Glucose Time Test Strip Symbol
Test Strip 1 Absorbent Hole Apply a drop of blood here. The blood will be automatically absorbed. 2 Confirmation Window This is where you confirm if enough blood has been applied to the absorbent hole in the strip. 3 Test Strip Handle Hold this part to insert the test strip into the slot. 4 Contact Bars Insert this end of the test strip into the meter. Push it in firmly until it will go no further. Front ATTENTION: The front side of test strip should face up when inserting test strip.
Features Model Data Transmission Function Test results can be viewed through a wireless connection by using 4G. For TD-3261G For TD-3261B 14 Test results can be viewed through a wireless connection by using Bluetooth.
SETTING THE MONITOR Before using your monitor for the first time or if you change the meter battery, check and update these settings. Make sure you complete the steps below and have your desired settings saved. ► Entering the setting mode Start with the monitor off (no test strip inserted). Press until the monitor turns on. Step 1 Setting the date. With the year flashing, press until the correct year appears. Press . With the month flashing, press until the correct month appears. Press .
With the day flashing, press until the correct day appears. Press . Step 2 Setting the time With the hour flashing, press until the correct hour appears. Press . With the minute flashing, press until the correct minute appears. Press . Step 3 Setting the time format Press to select the desired time format --- 12h or 24h. Press .
Step 4 Setting the measurement unit Press to select the desired blood glucose measurement unit- mg/dL or mmol/L. Press . Step 5 Setting a language Press to select L1 / L2. The default language for the meter is L1, which is English; L2 is French. To confirm your selection, press . Step 6 Setting the auto-send Press to select the auto-send On or OFF. Press . This function is referring to the 4G or Bluetooth service. If “On” is selected, your result will be transmitted right after the test.
Step 7 Deleting the memory While “dEL” and a flashing “ ” symbol appear on the display along with user numbers, if you do not wish to delete the saved results, press to skip this step. If you wish to delete ALL the results, press twice. Step 8 Setting the cycle mode Press to choose the desired blood pressure cycle mode from the following time interval selections: OFF, 5, 10, 15, 20, 30, 40 or 60 minutes. Press to turn off the meter.
TESTING YOUR BLOOD GLUCOSE The Four Measuring Modes The monitor provides you with four modes for measuring, General, AC, PC and QC. Modes Use when Gen any time of day without regard to time since last meal (not displayed) no food intake for at least 8 hours 2 hours after a meal testing with the control solution You can switch between each mode by: Step 1 Start with the monitor switched off. Insert a test strip to turn on the monitor.
Step 2 Press to switch between General, AC, PC and QC mode.
QUALITY CONTROL TESTING When should the control solution test be performed? • if it is mandatory following the local regulations in your country, • if you suspect the monitor or test strips are not working properly • if your blood glucose test results are not consistent with how you feel, or if you think the results are not accurate • to practice testing process, or • if you have dropped or think you may have damaged the monitor.
WARNING When doing the control solution test, you have to mark it so that the test result will NOT mix with the blood glucose TEST RESULTS be stored in the memory. Failure to do so will mix up the blood glucose test results with the control solution test results in memory. Step 3 Apply Control Solution Shake the control solution vial thoroughly before use. Squeeze out a drop and wipe it off, then squeeze another drop and place it on the tip of the vial cap.
Step 4 Read and Compare the Result After counting down to 0, the test result of control solution will appear on the display. Compare this result with the range printed on the test strip vial and it should fall within this range. If not, please read the instructions again and repeat the control solution test. Normal 109~165 Out-of-range results If you continue to have test results fall outside the range printed on the test strip vial, the monitor and strips may not be working properly.
TESTING WITH BLOOD SAMPLE NOTE To reduce the chance of infection: • Never share a lancet or the lancing device. • Always use a new, sterile lancet. Lancets are for single use only. • Avoid getting hand lotion, oils, dirt, or debris in or on the lancets and the lancing device. Preparing the Lancing Device for Blood Testing Please follow the instructions in the lancing device insert for collecting a blood sample.
• Rub the puncture site for about 20 seconds before penetration. • Clean the puncture site using cotton moistened with 70% alcohol and let it air dry. • Use a clear cap (included in the kit) while setting up the lancing device. ► Fingertip testing Press the lancing device’s tip firmly against the lower side of your fingertip. Press the release button to prick your finger, then a click indicates that the puncture is complete.
Alternative Site Testing IMPORTANT: There are limitations with AST (Alternative Site Testing). Please consult your health care professional before you perform AST. What is AST? Alternative site testing (AST) means that people use parts of the body other than the fingertips to check their blood glucose levels. This system allows you to test on the palm, the forearm or the upper arm, the calf or the thigh with the equivalent results to fingertip testing.
We strongly recommend that you perform AST ONLY at the following times: • In a pre-meal or fasting state (more than 2 hours since the last meal). • Two hours or more after taking insulin. • Two hours or more after exercise. Do NOT use AST if: • You think your blood glucose is low. • You are unaware of hypoglycemia. • You are testing for hyperglycemia. • Your AST results do not match the way you feel. • Your routine glucose results are often fluctuating.
Step 3 Obtaining a blood sample Use the pre-set lancing device to puncture your desired site. After penetration, discard the first drop of blood with a clean tissue or cotton. Gently squeeze the punctured area to obtain another drop of blood. Be careful NOT to smear the blood sample. The volume of blood sample must be at least 0.5 microliter (μL) of volume. ( actual size). Step 4 Apply the Sample Gently apply the drop of blood to the absorbent hole of the test strip at a tilted angle.
NOTE • Do not press the punctured site against the test strip or try to smear the blood. • If you do not apply a blood sample to the test strip within 3 minutes, the monitor will automatically turn off. You must remove and reinsert the test strip to start a new test. • The confirmation window should be filled with blood before the monitor begins to count down. NEVER try to add more blood to the test strip after your drop of blood has moved away. Discard the used test strip and retest with a new one.
TESTING YOUR BLOOD PRESSURE This monitor provides you two different ways to measure your blood pressure. Select from the options below: • Single measurement Perform an individual blood pressure measurement. • Average measurement Automatically perform three (3) consecutive blood pressure measurements and display the final average result. Before Measurement • Avoid caffeine, tea, alcohol and tobacco at least 30 minutes before measurement. • Wait 30 minutes after exercising or bathing before measurement.
► Fitting the Cuff Correctly Step 1 Connect the air plug of the tubing to the air jack of the monitor. Pile Side Material RED LINESHOULD BE AT THE INSID INDEX Step 2 Assemble the cuff as shown on the right. The smooth surface should be inside the cuff loop and the metal D-ring should not touch your skin. INDEX MUST Sewn Hook Material Step 3 Stretch your left (right) arm in front of you with your palm facing up.
RED LINESHOULD BE AT THE INSID INDEX Step 5 Press the hook material firmly against the pile material. The top and bottom edges of the cuff should be tightened evenly around your upper arm. INDEX MUST ►The range index of cuff should fall into this range. ► Proper Measurement Position Step 1 Sit down for at least 10 minutes before measuring. Step 2 Place your elbow on a flat surface. Relax your hand with the palm facing up.
Taking a Single Measurement Always apply the pressure cuff before turning on the monitor. Step 1 Press . All the LCD symbols will appear. Then the cuff will begin to inflate automatically. Step 2 The heart symbol “ ” will flash when a pulse is detected during the inflation. Step 3 After the measurement, the monitor displays the systolic pressure, diastolic pressure and pulse rate. Step 4 Press to switch off. Or it will switch off automatically after left idle for 3 minutes.
Average Mode Measurement Always apply the pressure cuff before turning on the monitor. Step 1 Press . The monitor will turn on and enter the averaging mode. Then the cuff will begin to inflate automatically. Step 2 After the first measurement is complete, the monitor will start counting down before the second measurement begins. The number on the right represents the remaining countdown between each measurement. The monitor will take three (3) measurements consecutively with an interval of 20 seconds.
Step 3 After taken three measurements, the results are averaged to produce the blood pressure measurement. Press to turn off the monitor. Auto Cycle Mode Always apply the pressure cuff before turning on the monitor. Step 1 Press to enter cycle mode. The monitor will automatically take blood pressure measurement in 2 minutes for first measurement. The CYC On will show on the display and keep still until the remaining time is in 1 minute.
Step 3 After completion of the cycle mode for 5 hours, the monitor will display “End” and press any key to turn off the monitor. NOTE To discontinue the cycle mode measurement during the time intervals, press and hold Press 36 for 3 seconds to enter setting mode and select cycle mode OFF. again to turn off the monitor.
MONITOR MEMORY Your monitor stores the 864 most recent test results along with respective dates and times in its memory. To recall the memory, start with the monitor off. Reviewing Test Results Step 1 . Press and release “ ”will appear on the display. Step 2 Press . Review all test results stored in the monitor. Press to search the result forwards, and press to search the result backwards. Step 3 Exit the memory. Press and the monitor will turn off.
Reviewing Day Average Results Step 1 Press and hold for 3 seconds. When “ ” appears on the display, keep pressing for 3 seconds until the flashing “ ” appears. Release and then your blood glucose or blood pressure 7-day average result will appear on the display. Step 2 Press . Review your 14- and 30-day average. Step 3 Press . Switch to review your blood glucose or blood pressure 7-, 14-, and 30-day average. Step 4 Exit the memory. Press and the monitor will turn off.
NOTE • Any time you wish to exit the memory, press or leave it without any action for 3 minutes. The monitor will turn off automatically. • Control solution results are NOT included in the day average. • If using the monitor for the first time, “---” displays when you recall the test results or review the average result. This indicates that there is no test result in the memory. • The blood glucose test results of and will NOT be calculated in the day average results.
DATA TRANSMISSION VIA BLUETOOTH (For TD-3261B) You can transmit your data from the meter to your devices via Bluetooth. Please contact your local customer service or place of purchase for assistance. Please note that you must complete the pairing between meter and Bluetooth receiver before transmitting data. Pairing with your mobile device 1. Turn on the Bluetooth function on your mobile device. 2. With the meter off. Press to turn Bluetooth on. “PCL” will appear on the display. 3.
NOTE • While the meter is in transmission mode, it will be unable to perform a test. • Make sure your device with iOS (7 or above) or Android System (4.3 or above) has turned on Bluetooth before transmitting the data and the meter is within the receiving range. • The Bluetooth functionality is implemented in different ways by the various mobile device manufacturers, the compatibility issue between your mobile device and the meter maybe occur.
DATA TRANSMISSION VIA 4G (For TD-3261G) For 4G Version Before initiating 4G transmission, make sure the SIM card and the battery are inserted correctly. 1. How to initiate 4G transmission Three methods can initiate 4G transmission (a) with the meter turned off, press to initiate 4G transmission of batch upload for current user with test results which have not been uploaded.
When the RF indicator is blinking in blue, the data transmission is waiting for connection. When the RF indicator is in solid blue, the 4G connection is established. Press and the meter will turn off. NOTE • If the SIM card is not inserted, the “PCL” will start blinking, as in the figure shown as right, and the meter will turn off automatically after blinking for 30 seconds. • If the monitor displays a message like the one on the right while connecting to the server, please try again.
The number displayed in the middle of the screen represents the signal quality level from 1 to 31. The greater the number, the better signal you have. We recommend that the signal quality should be above ten (10) when you perform the transmission. For the above 3rd image, the number displayed on the right represents the battery voltage value (38), if the number is below 37, it will be unable to initiate 4G transmission. Press to turn off the monitor when you finish.
MAINTENANCE Battery Your monitor comes with one 3.7V lithium-ion rechargeable battery. ► Low Battery Signal The monitor will display the two messages below to alert you when the monitor power is getting low. The “ ” symbol appears along with display messages: The monitor is functional and the result remains accurate, but it is time to recharge the battery. If the power is not enough to do a test, the symbol starts blinking. Please recharge the battery immediately.
NOTE • If the power indicator is blinking in red when connecting AC adapter to a power source, this indicates the monitor is in a pre-charging stage. Do not remove AC adapter power plug. The recharge will begin shortly. • If the battery is full-recharged but the power is getting low soon after use, please contact local customer service for assistance. Using AC Adapter ► Connect AC adapter to the monitor. Step 1 Connect AC adapter plug to AC adapter jack of the monitor.
CAUTION! The Li-ion battery must be recharged with the TD-3261G monitor by connecting AC adapter to an electrical outlet. Only this 3.7V rechargeable Liion battery (battery pack no. S14500 1S4P) can be recharged with TD-3261G by the connection of AC adapter. Any other kind of rechargeable battery is not allowed or it may damage the monitor. Caring for Your Monitor To avoid the meter and test strips attracting dirt, dust or other contaminants, please wash and dry your hands thoroughly before use.
Caring for Your Test Strips • Storage condition: 4°C to 40°C (39.2°F to 104°F) for strip vial and 2°C to 30°C (35.6°F to 86°F) for foil package, 10% to 85% relative humidity. Do not freeze. • Store your test strips in their original vial only. Do not transfer to other container. • Store test strip packages in a cool and dry place. Keep away from direct sunlight and heat. • After removing a test strip from the vial, immediately close the vial cap tightly. • Touch the test strip with clean and dry hands.
Important Control Solution Information • Use only our control solutions with your monitor. • Do not use the control solution beyond the expiry date or 3 months after first opening. Write the opening date on the control solution vial and discard the remaining solution after 3 months. • It is recommended that the control solution test should be done at room temperature (20°C to 25°C / 68°F to 77°F).
SYSTEM TROUBLESHOOTING If you follow the recommended action but the problem persists, or error messages other than the ones below appear, please call your local customer service. Do not attempt to repair by yourself and never try to disassemble the monitor under any circumstances. Result Readings MESSAGE WHAT IT MEANS BLOOD GLUCOSE Appears when your result is below measurement limit, which is less than 20 mg/dL (1.1 mmol/L). Appears when your result is between 20 to 69 mg/ dL (1.1 to 3.8 mmol/L).
Appears when your result is equal to or higher than 240 mg/dL (13.3 mmol/L). This indicates the possibility of ketone accumulation for type 1 diabetes. Please seek medical assistance immediately. Appears when your result is higher than the limit of measurement, which is higher than 600 mg/dL (33.3mmol/L). MESSAGE WHAT IT MEANS BLOOD PRESSURE Appears when your systolic pressure is below 140mmHg and your diastolic pressure is below 90mmHg.
Error Messages MESSAGE CAUSE Please contact local customer service for help. Problem with the measurement. Refit cuff tightly and correctly. Relax for 5 minutes and repeat the measurement. If the error still remains, please contact the local customer service for help. Problem with the meter. Please contact the local customer service for assistance. E-U Strip has been used. Repeat the test with a new strip. E-F Test strip is removed while counting down, or insufficient blood volume.
Error Messages of 4G Transmission MESSAGE CAUSE WHAT TO DO PCL No SIM card inserted. Insert your SIM card and place it correctly before transmitting data. PCL Err 27-30 PCL Err 40-45 PCL Err 50 PCL Err 62-69 Problem with upload. Please start over and try again. If the problem persists, please contact local customer service for assistance. PCL Err 31-37 PCL Err 72-76 Problem with setting. Please contact customer service for assistance. PCL Err 38 Unfinished registration.
Troubleshooting 1. Blood Glucose Measurement 1. If the monitor does not display a message after inserting a test strip: POSSIBLE CAUSE WHAT TO DO Test strip inserted upside down or Insert the test strip with contact incompletely. bars end first and facing up. Defective monitor or test strips. Please contact customer service. 2. If the test does not start after applying the sample: POSSIBLE CAUSE WHAT TO DO Repeat the test using a new test Insufficient blood sample.
3. If the control solution testing result is out of range: POSSIBLE CAUSE Error in performing the test. Control solution vial was poorly shaken. WHAT TO DO Read instructions thoroughly and repeat the test again. Shake the control solution vigorously and repeat the test again. Expired or contaminated control Check the expiration date of the solution. control solution. Control solution, monitor, and test Control solution that is too warm or strips should be at room too cold.
2. Blood Pressure Measurement 1. If nothing is displayed after pressing the On/Off button: POSSIBLE CAUSE WHAT TO DO Batteries incorrectly installed or Check that the batteries are absent. correctly installed. 2. If the heart rate is higher/lower than user’s average: POSSIBLE CAUSE WHAT TO DO Movement during measurement. Repeat measurement. Measurement taken just after Rest at least 30 minutes before exercise. repeating measurement. 3.
DETAILED INFORMATION Reference Values Blood Glucose The meter provides you with plasma equivalent results. Time of day Normal plasma glucose range for people without diabetes (mg/dL) Fasting and before meal Less than 100 mg/dL (5.6 mmol/L) 2 hours after meals Less than 140 mg/dL (7.8 mmol/L) Source: American Diabetes Association. Standards of Medical Care in Diabetes- 2018 Jan; 41(Supplement 1): S1-S2. Please consult your doctor to determine a target range that works best for you.
Definitions and Classification of blood pressure levels according to 2007 ESH-ESC Practice Guidelines for the Management of Arterial Hypertension: Category Systolic (mmHg) Diastolic (mmHg) Optimal < 120 and < 80 Normal 120–129 and/or 80–84 High normal 130–139 and/or 85–89 Grade 1 hypertension 140–159 and/or 90– 99 Grade 2 hypertension 160–179 and/or 100–109 Grade 3 hypertension ≥ 180 and/or ≥ 110 Isolated systolic hypertension ≥ 140 and < 90 Isolated systolic hypertension shou
SYMBOL INFORMATION Symbol Referent In vitro diagnostic medical device Consult instructions for use Type BF equipment Temperature limit Use-by date Batch code Manufacturer Serial number Caution Authorized representative in the European Community Humidity limitation CE mark This device does not belong to household waste and must be returned to a collection point for recycling electric and electronic devices according to local laws.
SPECIFICATIONS System performance Power source: Size of monitor w/o cuff: Memory: Power saving: One 3.6V rechargeable Li-ion battery 147 (L) x 105 mm (W) x 80 mm (H), 460g with batteries Maximum 864 memory records Automatic power off if system idle for 3 minutes (normal mode) or 5 minutes (PCL mode).
Blood glucose measurement performance Measurement unit: mg/dL or mmol/L Linear range: 20 to 600 mg/dL (1.1 to 33.3mmol/L) Precision: ±5 % (CV) ±15 mg/dL (0.83 mmol/L) when glucose < 100 mg/ Accuracy: dL (5.5 mmol/L) ±15% when glucose ≥ 100 mg/dL (5.5 mmol/L) Ketone warning: glucose value is above 240 mg/dL (13.
Warning: Medical electrical equipment needs special precautions regarding EMC and needs to be installed according to the EMC information provided. Careful consideration of this information is essential when stacking or collocating equipment and when routing cables and accessories. Warning: RF mobile communications equipment can effect medical electrical equipment.
Manufacturer’s declaration-electromagnetic emissions The TD-3261G is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the TD-3261G should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment-guidance (for home and professional healthcare environment) RF emissions CISPR 11 Group 1 The TD-3261G uses RF energy only for its internal function.
Manufacturer’s declaration-electromagnetic immunity The TD-3261G is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the TD-3261G should assure that it is used in such an environment.
Manufacturer’s declaration-electromagnetic immunity The TD-3261G is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the TD-3261G should assure that it is used in such and environment.
Recommended separation distance between portable and mobile RF communications equipment and the TD-3261G The TD-3261G is intended for use in an electromagnetic environment (for home and professional healthcare) in which radiated RF disturbances are controlled.
Manufacturer’s declaration-electromagnetic immunity Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment The TD-3261G is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the TD-3261G should assure that it is used in such an environment.
Federal Communications Commission (FCC) Statement 15.21 You are cautioned that changes or modifications not expressly approved by the part responsible for compliance could void the user’s authority to operate the equipment. 15.105(b) This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
FCC/IC RF Radiation Exposure Statement This equipment should be installed and operated with a minimum distance of 0 cm between the radiator and your extremity. This device and its antenna(s) must not be co‐located or operating in conjunction with any other antenna or transmitter. SAR is measured with the device at 0 mm to the extremity, while transmitting at the highest certified output power level in all frequency bands of the device.
TaiDoc Technology Corporation B1-7F, No. 127, Wugong 2nd Rd., Wugu Dist., 24888 New Taipei City, Taiwan www.taidoc.