User Manual

39
WARINING
It is extremely important that the cuff is at the same level of the heart. If the cuff
is relatively lower (higher) than the heart, the blood pressure measured will be
higher (lower) than the actual value. A 15 cm difference in height may result in
an error of 10 mmHg.
( Fig. 43)
(Fig. 44)
(Fig. 45)
NOTE
If you press the button during the measurement, the monitor will turn off.
39
STEP 6- All the LCD display symbols will
appear with a long “beep” sound (Fig. 43).
Then cuff begins to inflate automatically.
STEP 7- After the cuff pressure is reached,
the cuff begins to deflate. When detecting
pulse, the Pulse Symbol “ ” will flash,
along with a beeping sound during
measurement (Fig. 44).
STEP 8- After the measurement, the
monitor displays the systolic pressure,
diastolic pressure, and heart rate (Fig. 45).
Press the button to switch off. Or it will
switch off automatically after 3 minutes.