STEP 4- Apply Blood into the Test Strip
STEP 5- Obtain an accurate result in 7 seconds
The result of your blood glucose test is
shown after the meter counts to 0. This
reading is automatically stored in the
monitor (Fig.37).
PLEASE NOTE
1. Do not push your finger (with blood on it) against the test strip or try to apply
a smeared sample on the test strip.
2. If you do not apply blood sample within 3 minutes, or if you accidentally turn
off the monitor by pressing the button, the monitor will alert you with
flashing test strip symbol and continuous beeping. You will need to turn the
monitor off by removing the test strip.
3. The blood should be completely filled the confirmation window before the
monitor begins to count down. If you find that the confirmation window is not
filled with blood when the monitor is counting, NEVER try to add more blood
to the test strip. Discard the test strip and retest with a new one.
4. To start a blood pressure test after performing a blood glucose test, make
sure that the test strip is removed from the monitor. Then turn the monitor
off and refer to "HOW TO MEASURE BLOOD PRESSURE".
33
When “ “is flashing on the screen, apply your
blood to the absorbent hole of the test strip until
the confirmation window is fully covered with
blood(Fig. 36). The monitor then begins to count
down automatically.
(Fig.36)