User Manual
26
Doing a Control Solution Test
TAKE A TEST STRIP OUT WITH CLEAN AND DRY HANDS FIRST.
.
Be sure the code number on the display is the same as the code number on
the test strip vial. If the code numbers are not identical, please see “Coding the
Monitor” section of this manual.
NOTE
Contact bars must be inserted all the way into the meter or you may get an
inaccurate test result.
Every time you perform a control solution test, you must enter into the “CTL”
test mode so that the test result will not be stored in the monitor memory.
Failure to do so will confuse the blood glucose test result with the control
solution test result in memory.
26
(Fig.20)
STEP1- Insert Test Strip
Insert a test strip with contact bars end first and facing
up,into the test slot.The monitor turns on automatically and
displays the followings in sequence:
“CHK” and “ ” Îambient temperatureÎcode number
and flashing“ ”(Fig.20)
STEP 2- Press the M button
While the “ ” symbol appears on the
display, press the M button (Fig.21) and
then “CTL” will appear on the display
(Fig.22). With the “CTL” sign on the
display, the meter will not store your test
result in memory. If you decide not to
perform a control solution test, press the
M button again, and the “CTL” sign will
disappear. (Fig.22) (Fig.21)