Important Control Solution Information
Use only TaiDoc control solutions.
Check the expiry date on the control solution vial. Do not use if expired.
Control solution, monitor, and test strips should come to room temperature
(68-77℉/20-25℃) before testing.
Shake the vial, discard the first drop of control solution, and wipe off the
dispenser tip to ensure a good sample and an accurate result.
Use only for 90 days after first opening. Record the discard date (date
opened plus 90 days) on the control solution vial. Discard after 90 days.
Store the control solution tightly closed at temperatures below 86℉ (30℃).
Do not refrigerate.
NOTE
The control solution range printed on the test strip vial is for Taidoc control
solution only. It is used to test monitor and test strip performance. It is not
recommended range for your blood glucose level.