Owner's Manual
A2.1‐28of30
SPECIFICATIONS
System performance
Power source: Four 1.5V AAA alkaline batteries
Size of meter w/o cuff:
145 (L) x 64.5 (W) x 29.6 (H)mm,
240g without batteries.
Power saving: Automatic power off if system idle for 3 minutes.
System operating condition: 5℃ to 40℃ (41℉ to 104℉), 15% to 93% RH,
700 hPa to 1060 hPa
Meter storage condition: -25℃ to 70℃ (-13℉ to 158℉), 10% to 95% RH
Waterproof grade IP22
The expected service life 2 years
Blood pressure measurement performance
Pressure range: 0 - 300 mmHg
Heart rate range: 40 - 199 beat per minute
Systolic Measurement Range: 50 mmHg -250 mmHg
Diastolic Measurement Range: 30 mmHg -180 mmHg
Pulse Rate Measurement Range: 40 -199 beats / minute
Maximum inflation pressure: 280 mmHg
Accuracy of Pressure: ±3 mmHg or ±2% of reading
Accuracy of Pulse rate: ±4% of reading
Measurement unit: mmHg
This device has been tested to meet the electrical and safety requirements of: IEC 60601-1,
and IEC 60601-1-2
Reference to Standards:
• IEC60601-1 General requirement for safety
• IEC60601-1-2 Requirements for EMC
IP22:
• Object size protected against: >12.5 mm (Fingers or similar objects)
• Protected against: Dripping water when tilted up to 15°