Owner's Manual

A2.1‐12of30
3. Leave a little free space between the arm and the cuff; you should be able to
fit two fingers between them. Clothing must not restrict the arm. Remove all
clothing covering or constricting the measurement arm.
4. Press the hook material firmly against the pile material. The top and bottom
edges of the cuff should be tightened evenly around your upper arm.
Proper Measurement Position
1. Sit down for at least 10 minutes before measuring.
2. Place your elbow on a flat surface. Relax
your hand with the palm facing up.
3. Make sure the cuff is about the same height
as the location of your heart.
4. Press ON/OFF Button. Remain still and do not talk or move during the
measurement.
WARNING:
If the cuff is relatively lower (higher) than the heart, the obtained blood
pressure value could be higher (lower) than the actual value. A 15 cm
difference in height may result in an error around 10 mmHg.
5.
Measurement is in progress.
After the meter is turned on, the cuff will begin to inflate automatically.