Fora Care Blood Pressure Monitoring System Model number: P80 Owner’s Manual Version 1.0 2013-06 A2.
Dear FORA Blood Pressure Monitoring System Owner: Thank you for purchasing the FORA Blood Pressure Monitoring System. This manual provides important information to help you operate this system correctly. Before using this product, please read the following contents thoroughly and carefully. With the compact size and easy operation of this system, you can easily monitor your blood pressure by yourself at any time or place.
IMPORTANT SAFETY PRECAUTIONS READ BEFORE USE 1. Use this device ONLY for the intended use described in this manual. 2. Do NOT use the device if it is not working properly or damaged. 3. Do NOT use the equipment where aerosol sprays are being used, or where oxygen is being administered. 4. Do NOT use under any circumstances on newborns. 5. This device does NOT serve as a cure for any symptoms or diseases. The data measured are for reference only. Always consult your doctor to have the results interpreted. 6.
TABLE OF CONTENTS BEFORE YOU BEGIN Intended Use Test Principle Contents of System Meter Overview Display Screen HOW TO INSTALL AND UPDATE THE TDLink BP BLUETOOTH PAIRING IMPORTANT INFORMATION ABOUT PERFORMING YOUR BLOOD PRESSURE TEST Before Measurement Fitting the Cuff Properly Proper Measurement Position HOW TO TEST YOUR BLOOD PRESSURE Taking Measurement by the Device Itself Taking Measurement with the TDLink BP App VIEWING PAST TEST RESULTS Data Analysis FAQ MAINTENANCE Battery Caring for Your Meter DETAI
BEFORE YOU BEGIN Intended Use FORA Blood Pressure Monitoring System is a system designed to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. It is intended for individuals age 18 and above use at home and in clinical settings. The device is not to be used for the diagnosis of hypertension or for testing on newborns.
Meter Overview 1. 3. 2. 4. 6. 5. 1. CUFF 2. BT INDICATOR Download test results with a Bluetooth connection. 3. DISPLAY SCREEN 4. BATTERY COMPARTMENT 5. ON/OFF BUTTON 6. BT PAIRING BUTTON A2.
Display Screen 3. 2. 1. 4. 5. 8. 7. 1. Systolic Pressure Value 2. Diastolic Pressure Value 3. Pulse Rate Symbol 4. Low Battery Symbol 6. 5. Pulse Rate 6. Diastolic Pressure Symbol 7. Systolic Pressure Symbol 8. Units of Blood Pressure A2.
HOW TO INSTALL AND UPDATE THE TDLink BP . You can use the device, FORA Blood Pressure Monitoring System, by itself or with the TDLink BP App. TDLink BP App is available on the App Store/Play Store for free download. You must be connected to the internet to download the App. TDLink BP App is designed to assist you in easy blood pressure testing, recording, tracking and monitoring. It is simple and intuitive to use, for better understanding of your current condition and to achieve better hypertension control.
BLUETOOTH PAIRING Please note that you must complete the pairing between meter and Bluetooth receiver before transmitting data. NOTE: This step is recommended when the user needs to pair this meter to a Bluetooth receiver for the first time, or when user needs to pair this meter to another new Bluetooth receiver. 1. Installing the TDLink BP app For more information, please refer to the section” HOW TO INSTALL AND UPDATE THE TDLink BP”. 2.
Bluetooth indicator on the blood pressure meter: Bluetooth Indicator Flash fast Flash slowly Lit solid Status The meter is searching the device of Bluetooth signal. The meter is pairing with the device of Bluetooth. The meter is transmitting the data now the connection is established. Mobile Phone Compatibility Issues The Bluetooth functionality is implemented in different ways by the various mobile phone manufacturers.
IMPORTANT INFORMATION ABOUT PERFORMING YOUR BLOOD PRESSURE TEST Before Measurement Avoid caffeine, tea, alcohol and tobacco for at least 30 minutes before measurement. Wait 30 minutes after exercising or bathing before measurement. Sit or lie down for at least 10 minutes before measurement. Do not measure when feeling anxious or tense. Take a 5-10 minute break between measurements.
3. Leave a little free space between the arm and the cuff; you should be able to fit two fingers between them. Clothing must not restrict the arm. Remove all clothing covering or constricting the measurement arm. 4. Press the hook material firmly against the pile material. The top and bottom edges of the cuff should be tightened evenly around your upper arm. Proper Measurement Position 1. Sit down for at least 10 minutes before measuring. 2. Place your elbow on a flat surface.
HOW TO TEST YOUR BLOOD PRESSURE Taking Measurement by the Device Itself Always apply the pressure cuff before turning on the meter. 1. Press ON/OFF Button. All the LCD symbols will appear with a long “beeping” sound. Then the cuff will begin to inflate automatically. 2. Pump starts to inflate and pressure in cuff increases to working pressure. 3. The heart symbol “ ” will show silently when pulse is detected during the deflation. When signal of pulse becomes strong, “ ” icon starts to flash. 4.
Taking Measurement with the TDLink BP App 1. Turn on the Bluetooth switch on your mobile and connect with “FORA Blood Pressure Monitoring System” before doing the test. 2. Turn on TDLink BP App; always apply the pressure cuff before turning on the meter. 3. Press ON/OFF Button. All the LCD symbols will appear with a long “beeping” sound on the device monitor. Then the cuff will begin to inflate automatically until it reaches the working pressure. 4. Pulse rate is detected during the deflation. 5.
6. Press ON/OFF Button to switch off. Or it will switch off automatically after being idle for 3 minutes. NOTE If you press ON/OFF Button during measurement, the meter will be turned off. A2.
VIEWING PAST TEST RESULTS Data Reviewing Data Record The meter stores the blood pressure test results along with respective dates and times in memory under the mobile device. Data Record Screen Blood Pressure Test Result Blood Pressure unit Pulse Rate Test Result Pulse Rate Unit 1. Press “Data” on the main menu bar to review the records. The first reading you see is the last blood pressure result. 2. Swipe left or right to recall the previous test results stored in the memory. A2.
VIEWING PAST TEST RESULTS Analysis Reviewing Blood Pressure Day Average Results The TDLink BP App offers you 7-, 14-, and 30- day average results and Pie Chart to assist you to monitoring your blood pressure. Day Average Screen 7-Days of Blood Pressure Control Ratio in Percentage Blood Pressure Day AVG Result Detail Information Distribution of 14 Days Blood Pressure Results Pulse Rate Day AVG Result 1. Press “Analysis” on the main menu bar to enter the Day AVG records. 2.
A2.
FAQ 1. Select the FAQ mode from the Main Menu Bar. Select FAQ in the Main Menu Bar to enter, and then the Sub Menu will appear. You will see the information categorized into different sections, such as Self-Monitoring, Diet Control, Exercise, Medication and Overview. If you wish to find the specific information much quicker, you can also search by typing key words in the search bar on the top of each section page. A2.
MAINTENANCE Battery Your meter comes with four (4)1.5V AAA size alkaline batteries. Low Battery Signal The meter will display either of the two messages below to alert you when the meter power is getting low. 1. The symbol appears along with display messages: The meter is functional and the result remains accurate, but it is time to change the batteries. 2. The symbol appears with E-b: The power is not enough to do a test. You must change the batteries immediately. A2.
Replacing the Battery To replace the batteries, make sure the meter is turned off. 1. Press the edge of the battery cover and lift it up to remove. 2. Remove the old batteries and replace with four 1.5V AAA size alkaline batteries. 3. Close the battery cover. NOTE: As with all small batteries, these batteries should be kept away from small children. If swallowed, promptly seek medical assistance. Batteries might leak chemicals if unused for a long time.
Caring for Your Meter To avoid the meter attracting dirt, dust or other contaminants, wash and dry your hands thoroughly before use. Cleaning To clean the meter exterior, wipe it with a cloth moistened with tap water or a mild cleaning agent, then dry the device with a soft dry cloth. Do NOT flush with water. Do NOT use organic solvents to clean the meter. Do NOT wash the pressure cuff. Do NOT iron the pressure cuff.
DETAILED INFORMATION Reference Values Clinical studies show that adult diabetes is often accompanied by elevated blood pressure. People with diabetes can reduce their heart risk by managing their blood pressure along with diabetes treatment*4. Monitoring your routine blood pressure trend helps you to know your body condition. Human blood pressure naturally increases after reaching middle age. This symptom is a result of continuous ageing of the blood vessels.
*4: American Diabetes Association: The Diabetes-Heart Disease Link Surveying Attitudes, A2.
SYSTEM TROUBLESHOOTING If you follow the recommended action but the problem persists, or error messages other than the ones below appear, please call your local customer service. Do not attempt to repair by yourself and never try to disassemble the meter under any circumstances. Result Readings of App MESSAGE WHAT IT MEANS Reading in yellow when blood pressure is low. Reading in green when blood pressure is normal. Reading in red when blood pressure is high. A2.
Error Messages of Meter MESSAGE CAUSE WHAT TO DO Inflation or pressure error Please contact local customer service for help. Refit cuff tightly and correctly. Relax Blood pressure measurement error. and repeat the measurement. If error still remains, contact local customer service for help. Review the instructions and repeat the test. If the meter still does not Problems with the meter. work, please contact the local customer service for help. Battery is too low. Repeat with new batteries. A2.
Troubleshooting 1. If no display appears after pressing ON/OFF Button: POSSIBLE CAUSE WHAT TO DO Batteries exhausted. Replace the batteries. Batteries incorrectly installed or Check that the batteries are correctly absent. installed. 2. If the heart rate is higher/lower than user’s average: POSSIBLE CAUSE WHAT TO DO Movement during measurement. Repeat measurement. Measurement taken just after exercise. Rest at least 30 minutes before repeating measurement. 3.
SPECIFICATIONS System performance Power source: Four 1.5V AAA alkaline batteries Size of meter w/o cuff: 145 (L) x 64.5 (W) x 29.6 (H)mm, 240g without batteries. Power saving: Automatic power off if system idle for 3 minutes.
FEDERAL COMMUNICATIONS COMMISION (FCC) STATEMENT 15.21 You are cautioned that changes or modifications not expressly approved by the part responsible for compliance could void the user’s authority to operate the equipment. 15.105(b) Federal Communications Commission (FCC) Statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules.
operating instructions for satisfying RF exposure compliance. Distributed by: Fora Care Inc. 893 Patriot Dr., Suite D Moorpark, CA 93021 U.S.A. Toll Free: 1-866-469-2632 Hours of Operation: 24 hours, 7 days/week A2.