User Instructions

T12-Assistent, Version 1.2
T12CVS8 HBM: public 65
Important
When a rotor and stator are replaced and the ID is different, the calibration
setting is automatically deactivated!
The same thing happens if the output signal scaling of the frequency output is
changed in the T12(HP). The calibration setting can and must only be used in
the status of scaling that was set during the calibration of the transducer
(normally the factory setting).
Information
If the Calibration active OFF setting is not saved with "Save to transducer" and
the original rotor is reinstalled, the calibration table (after restarting T12HP with
OFF / ON) automatically becomes active again
Deleting/ removing a calibration
Calibrations of frequency / voltage are saved in the respective parameter sets.
4 pairs are possible with 4 parameter sets. To remove these from a parameter
set, "Delete/remove calibration" must be set to ON and confirmed with OK.
Then save the parameter set (this is necessary to delete the calibration from
memory!). Otherwise the original calibration will not be active again after a
restart.
The unit in the calibration window is determined at the time when the
calibration values are saved by the unit that is currently valid. It remains in
effect if the unit changes, but the calibration becomes inactive.
Copying and saving calibrations
A calibration can be copied from one parameter set to another with "Save to
parameter set"