User Instructions

T12-Assistent, Version 1.2
64 HBM: public T12CVS8
Settings compatible
The "Compatible" entry is read only and cannot be changed. It indicates that
calibration can be activated (ON) because important parameters have not been
changed (scaling, frequency range, rotor and stator ID, unit)
To overwrite/change values, the Write/overwrite calibration field must be set to
ON (which serves as protection against accidental overwriting) otherwise the
changes will not be applied. This does not apply to the Calibration active ON/
OFF item.
Writing / overwriting a calibration
The date, calibrator's initials and cycle are linked with this information and form
a unit.
Also the rotor and stator ID and the unit of measure.
When rotors and stators with a different ID are used, for example when a rotor
or stator is replaced, calibration is automatically deactivated (set to OFF) and
therefore switched to invalid. Then the rated output (nominal) is automatically
used for scaling
Important
For a deactivated calibration in TEDS to be represented as current, the stator
must be restarted (which is generally done when a rotor is exchanged) or the
current status/parameter set must be resaved ("Save to transducer" is linked
with write to TEDS). Without a restart or "Save to transducer", modified
settings/states are not updated in the Assistant in the TEDS data. This applies
to all settings.