User Instructions

T12-Assistent, Version 1.2
T12CVS8 HBM: public 63
with the TEDS editor (for example QuantumX Assistant) and the entries can
be checked.
Important
Any change to the TEDS information (calibration) must be confirmed with 3)
OK and confirmed to save in the parameter set with 5). Otherwise it will not be
active.
Important
The TEDS of the T12HP may only be edited / written to with the T12 Assistant!
Writing with the TEDS editor, for example, is not permitted. The T12 Assistant
is always the TEDS master. If TEDS is written to with the QuantumX TEDS
editor despite this, the next time the T12/T12HP restarts or when the
parameter sets are loaded, the content will be overwritten by the valid values
(parameter set) of the T12 Assistant.
Information
The latest Assistant software for QuantumX and the TEDS editor (DLL) must
be used to detect T12HP.
Calibration active ON/OFF is used to control whether the calibration table will
be written to TEDS during the TEDS write process (restart or write parameters
menu). The downstream QuantumX amplifier series MX460B or MX840B/440B
is then able to use the calibration data saved in TEDS for compensation.
Normally information is always written to the measuring point.
The calibration points are the two points that are entered while calibrating on
the calibration machine. They indicate the calibration protocol. For calibration
to DIN51309 or VDI2646 we recommend using the rated output derived from
the specified interpolation equation(s).