50 HBM: public T12CVS8
Contents of the TEDS memory as defined in IEEE 1451.4
The information in the TEDS memory is organized into areas, which are
prestructured to store defined groups of data in table form.
Only the entered values are stored in the TEDS memory itself. The amplifier
firmware assigns the interpretation of the respective numerical values. This
places a very low demand on the TEDS memory. The memory content is
divided into three areas:
Area 1
An internationally unique TEDS identification number (cannot be changed).
Area 2
The base area (basic TEDS), to the configuration defined in standard IEEE
1451.4. The transducer type, the manufacturer and the transducer serial
number are contained here.
Example: TEDS content of a T12HP/1 kN·m transducer
TEDS
Manufacturer
HBM (31)
Model
T12HP (15)
Version letter
A
Version number
2 first position of stator ID no.
Serial number
7 first position of stator ID no.
Area 3
Data specified by the manufacturer and the user are contained in this area.
Typical values for an HBM T12HP/1 kN·m torque transducer are shown in the
"Value" column of the tables below.
Torque
HBM has already writte
”
n the “Frequency/Pulse Sensor” and “High Level
Voltage Output Sensor templates for the measured quantity of torque.