User Instructions

Load-carrying capacity
46 HBM: public T12CVS8
11
Shunt signal
The T12CVS8 torque transducer delivers a shunt signal, at either 50% or 10%
of the nominal (rated) torque, as selected. Activate this function via the T12
Assistant or the shunt signal resolution on plug 1 or plug 3 (see Section 10.3).
The shunt signal most recently selected in the T12 Assistant is then trig-
gered.
Information
Internal signal conditioning can cause a delay in triggering of about 5 seconds.
To achieve stable conditions, it is advisable to only activate the shunt signal
once the transducer has been warming up for 15 minutes.
The boundary conditions for reproducibility (e.g. the mounting conditions) must
be established in order to reproduce the measured values in the test report.
Important
The transducer should not be under load when the shunt signal is being mea-
sured, as the signal is applied additively.
Information
After about 5 minutes, the shunt signal is automatically deactivated.