Electrical connection
38 HBM: public T12CVS8
In the case of interference due to potential differences (compensating
currents), supply voltage zero and housing ground must be disconnected on
the amplifier and a potential equalization line established between the stator
housing and the amplifier housing (copper conductor, 10 mm
2
wire cross-
section).
If potential differences arise between the rotor and the stator on the machine,
perhaps due to unchecked leakage, and this causes interference, it can usually
be overcome by connecting the rotor directly to ground by a wire loop, for in-
stance. The stator should be fully grounded in the same way.