User Instructions
Electrical connection
36 HBM: public T12CVS8
10.1.1
Using the EMI filter/toroidal strip-wound core, for applications in
the USA and Europe
To suppress high frequencies, use an EMI filter/toroidal strip-wound core for
the cable through which the transducer is supplied. For use in the USA and in
the EU, work with 5 cable windings for the 100N·m and 200N·m measurement
ranges. Apply 3 cable windings for all other measurement ranges.
The mounting must be implemented with cable ties that are suitable for the
particular application. Choose an area for the mounting that is not exposed to
any mechanical stress (i.e. no unwelcome vibrations, etc.).
Fig. 10.1 EMI filter/toroidal strip-wound core installation example
Information
Take into account approx. 40 cm (3 windings), or approx. 70 cm (5 windings) of
additional cable length to install the EMI filter/toroidal strip-wound core.
Installation mounting
EMI filter/
toroidal strip-wound core
Cable tie
5 windings for 100Nm, 200Nm
3 windings for all other
measurement ranges