User's Manual

3
Doc. #.: PR2QM-01_rev.E00(20.09.23)
product.
2. Press the Power/Measure button to turn on
the power.
Measuring body temperature
1. Press the Mode setting button to select out of
“People, measurement sound ON” / “People,
measuring sound OFF” / “Object,
measurement sound ON” for setting.
2. Place it at the measurement area (temporal
artery) then press the Measure button.
3. The measured temperature can be checked in
the measurement display and the high
temperature/low temperature status is
indicated in the LED display light.
4. When the body temperature is measured, the
surrounding temperature and humidity are
measured together and they can be checked
in the Mobile App.
1) Measurement range below 36.0°C: Orange
light
2) Measurement range of 36.0°C ~ 37.6°C:
Green
3) Measurement range of 37.6°C ~ 38.5°C:
Orange
4) Measurement range exceeding 38.5°C: Red
Measuring object temperature
1. Press the Mode setting button to set as
“Object, measurement sound ON”.
2. After placing it at the object subject to
measurement, press the Measure button then
release.
3. The measured temperature can be checked in
the measurement display.
4. The temperature measurement range of object
is 10°C ~70°C.
Measuring surrounding temperature /
humidity
1. Press the Mode setting button for more than3
seconds to measure the surrounding
temperature/humidity.
2. After the surrounding humidity is displayed for
3 seconds, the surrounding temperature is
displayed.
(※ Information on the surrounding temperature
and humidity that have been measured must
only be checked for reference.)
Precautions before use
1. Learn how to use the product by reading the
User Manual thoroughly.
2. Use in surrounding temperature of 16°C ~
40°C. (Use after it has been indoors for 30
minutes).
3. Maintain an appropriate distance from direct
fire that could interfere with the measurement.
4. Check the connection status of the battery.
5. Check if there is damage or abnormality in the
device.
6. Be aware of damage caused by impact.
7. The body temperature may differ slightly
(0.1~0.4°C) depending on the measurement
area (mouth, ear, forehead).
8. To save the battery, it is better to connect to
THERMOCARE mobile App through Bluetooth
after several measurements.
Operating condition and storage
environment
Operating temperature: 16 ~ 40
Storage temperature and humidity: -20 ~
50 /95% below
How to store and manage after use
1. Store away from water.
2. In case it is not used for a long period of time,
remove the batteries.
3. Keep it safety, away from vibration, impact and
the like.
4. Store away from children under the age of 12.
5. Avoid direct fire and sunlight. Store in places
without dust or pollutant.
6. This product does not require inspection,
calibration or maintenance during use.
However, if an error in body temperature
measurement occurs or is expected to occur,
request the manufacturer for correction or
inspection.
7. Discretionary disassembly of the device and
replacement of internal parts are not
permitted. All information on repair and
replacement of the device (part replacement
method and circuit diagram, etc.) is provided
to our A/S team and must be performed by
the personnel designated by our company.
In case damage has been discovered, do not
attempt to repair. Contact the Customer
Center.
8. Keep in room temperature and if it has not
been stored in room temperature, store the
product in room temperature for 30 minutes
before using the product.
9. The probe (temperature measuring part) of the
thermometer is an extremely sensitive part so
it must be clean and free from damage at all
times for accurate measurement. Maintain a
clean status and store in dry places.
Precautions during use
1. The temperature measured at the right side of
the forehead and at the left side of the
forehead may differ.
2. Do not use this device for unintended purpose.
3. Measurement may be inaccurate in case there
is insufficient battery. In such a case, change
the battery before use.
4. Ensure that babies and infants do not suck the
probe (temperature measuring part) of the
thermometer or the battery or touch with wet
hands.
5. In case the baby or infant has come indoors
from outside or in case the thermometer has
been placed in a place with different
temperature environment, use it after 30
minutes.
6. In case there is sweat on the forehead,
measure the area behind the earlobe instead
of the forehead. Lift the hair around the ears
completely and measure the area behind the
earlobe.
7. Measure repeatedly for more than 3 times in