User's Guide

Etape 3 : Déterminer la concentration de Trichloramine Page 17/21
User guide of the Analysis kit for Trichloramine on the air
4) Error messages
"Error: trichloramine concentration below the measurement scale" :
This message may appear if:
- the measurement made at a trichloramine level which is too low and is not within the
minimum measuring range that can be measured
- the analysis cuvette is poorly positioned in the trikloram and contacts with the electrodes are
not well done
o Reposition the cuvette, the cork must fit perfectly into the intended housing
"Error: Trichloramine concentration above the measurement scale":
This message may appear if:
- the measurement carried out has a trichloramine level that is too high and is not within the
maximum measuring range that can be measured
- the analysis tank has been polluted
o Clean the analysis tank by rinsing 3 times with deionized water
For any other value, contact the SYCLOPE Electronique technical department