User's Guide
Etape 2 : Prélever un échantillon d’air Page 8/21
User guide of the Analysis kit for Trichloramine on the air
V. Step 1 : Cleaning the equipment
1) Equipment to clean
Id Designation
3 Cuvette of analysis
4 Filter clamp
5 Syringe 10 mL
2) Cleaning the filter clamp
N° Action Picture
1
Clean the filter clamp [4] by pouring
demineralized water over it.
3) Cleaning the syringe
N° Action Picture
1
Clean the outside of the syringe [5] by pouring
demineralized water over it.
2
Clean the inside of the syringe [5] by filling it
and emptying it several times with
demineralized water.
This step must be done :
- Before each new sample
Cleaning must be done with demineralized water, this step is essential before sampling.
All utensils in contact with the sample as well as the test solution must be sound.
Improper cleaning will lead to a flawed analysis!