Installation Instructions
Calibration Page 17/24
Installation and commissioning instructions
VII. Calibration
CAA320X et CAA330X probes require a calibration of the slope of the sensor and, if necessary (low
chlorine measurement), a calibration of the zero point. Depending on the physio-chemical parameters
of the water, the gain multiplier switch can be used.
1) Calibration of the sensor slope
A calibration of the slope of the sensor is required after the first commissioning (about 3h) and on a
regular basis for normal chlorination operation. If chlorine dioxide is present, the calibration shall take
account of its presence. The action of chlorine dioxide is 6 times greater than the measured chlorine.
Circulate chlorinated water in the chamber, wait for stabilization of the reading on the
control device.
The chlorine value must be at least 10% of the measurement scale.
Perform a chlorine measurement using a reference device.
Perform a countermeasure to validate this standard value.
Enter this value in the regulator to validate calibration.
In order to perform a correct calibration, the probe must be used in the measuring chamber,
at nominal pressure with the recommended flow rate (see specifications).
2) Zero-point calibration (if required)
Zero-point calibration is required when measuring at low chlorine value. In principle, the cell has a very
stable “zero” in absence of chlorine or any other oxidant. However, in order to ensure a perfect zero
integrating the drifts of the measuring chain, it should be verified and, if necessary, calibrated.
Clear the calibrations on the analyzer.
Turn off water circulation in the chamber.
Allow values to stabilize for several minutes.
Once the stability of the measure is perfect, make the zero on the analyser
Reopen the circulation of water
After the zero-point calibration a calibration of the slope of the probe should be done (See
§ above)
3) Use of the gain multiplier switch
Depending on the pysico-chemical parameters of the water, the gain multiplier switch can
be used. This switch allows to multiply the gain of the slope by 5.
Example 1 :
The use of stabilizer will cause a drop in chlorine measurement. If necessary, use
the gain multiplier switch by passing it in position X5.
Example 2 :
In high pH seawater the measurement of chlorine or brome will be low. If
necessary, use the gain multiplier switch by passing it in position X5.
X5
X1
X5
X1
X5
X1