Programming Instructions

Programming screen « INSTALLATION » Page 34/74
Fault Low range
- Selection of the current that will be applied on the output in case of fault encountered, on this
selection button we find the selected current information (
Low range).
Press to change it.
Low range = 4mA if range 4…20mA selected; 0mA if range 0…20mA selected.
0mA = 0mA
2,6mA = 2,6mA
Out of range 20mA
- Selection of the current that will be applied on the output in case of exceeding the selected operating
range, on this selection button we find the selected current information (
20mA).
Press to change it.
Maximum = higher 22mA
0mA = 20mA
20,8mA = 20,8mA
To delete a 4…20mA output :
- You must select « NO » in the 4…20mA output function and save your configuration by pressing the
“SAVE” button.
6) Menu « Configuration & Installation » - « CONDITIONS » [0631]
« CONDITIONS » menu will allow you to access the configuration of the operating conditions that
are used to control the parameters operation.
Press to open the following screen.
The screen displays the list of configurated and conditionable parameters. You can, by pressing a used
parameter, change the configuration of the applied conditions.
Type of parameter conditionable.
Press to edit the configuration.
Parameter type inactive and not conditionable.
CONDITIONS
Conditions setting
Conditional parameters
pH Free chlorine
Temperature
Not used
Not used
Not used
Not used
Not used