6.3.2 Non-centre installation
In bigger pipe diameter (> DN150) situations, the non-centre
installation is applicable. In the non-centre installation, the sensor is
inserted to exact 100 mm-depth of pipes.
Insertion depth= x+ y+ 100
x is the wall thickness of pipes
Example for a 12” pipe, the wall thickness of pipes 9 mm and a 87 mm
ball valve:
x= 9 mm ; y=87mm
Insertion depth=9 mm+ 87 mm+ 100 mm=196 mm