Operation Manual

9. Sensor signal outputs
9. Sensor signal outputs
9.1 Analog output
The sensor has an analog output range of 4 ... 20 mA. This output can be
scaled to match a desired measuring range. Standard scaling is from 0 to max
flow. The corresponding flow in different pipe sizes can be found in Appendix A.
For other ranges, please contact the manufacturer.
9.2 Pulse output
The sensor sends out one pulse per consumption unit. This pulse output can be
connected to an external pulse counter to count the total consumption. The
number of m
3
per second is summed up and indicated after one second. Pulse
length depends on flow rate.
S 421 21