Operation Manual

9. Sensor signal outputs
In case the flow rate is too high the S401 can not output the pulses with
default settings (one pulse per consumption unit). For this the pulse can
be set by our service software or a connected display to 1 pulse per 10
consumption units or 1 pulse per 100 consumption units. For example,
if set to 1 pulse per 10 m
3
, the sensor will send one pulse each 10 m
3
.
Example (1 pulse per 10 m
3
):
Volumetric flow
[m
3
/s]
Volumetric flow
[m
3
/h]
Pulse length
[ms]
Max. pulse
output
per hour
3 10800 120 1080
> 3 > 10800 60 2880
> 6 > 21600 30 3960
S401 23