Instruction Manual

2.5 Environmental Specifications Temperature
Temperature
Operating: +50~+104°F / +10~ +40°C
Storage/Transportation: -4~+140°F / -20~+60°C
Humidity
Operating: 15~95%, noncondensing
Storage/Transportation: 10~95%, noncondensing
Atmosphere Pressure
Operating: 70~106kpa
Storage/Transportation: 50~107.4kpa
2.6 Display
Display Color: 1.3”,Blue
Display content: SpO2%, Pulse Rate, PI%, Bar Graph Battery Indicator,
Pulse Wave
Notes:
1. The claim for oxygen saturation accuracy should be supported by
clinical studies covering the entire claimed range. The fraction of
inspired oxygen (FiO2) delivered to test subjects is varied to achieve
a series of targeted steady-state saturation periods over the specified
SpO2 accuracy range (e.g. 70 % to 100 %),then the SpO2 accuracy is
calculated by comparing SpO2 readings of the pulse oximeter to the
values of SpO2 determined with a Co-Oximeter.
2. The clinical trial included 11 subjects, including 6 males and 5 fe-
males, with an age range of 18 to 46 years and with varied skin color.