Instruction Manual

Guidance and manufacturer’s declaration – electromagnetic immunity
The RES5102 is intended for use in the electromagnetic environment
specified below. The customer or the user of the RES5102 should
assure that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Com-
pliance
level
Electromagnetic environ-
ment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms 150 kHz
to 80 MHz
6 Vrms 150
kHz to 80 MHz
outside ISM
bandsa
N/A Portable and mobile
RF communications
equipment should be
used no closer to any part
of the Model RES5102,
including cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7
GHz
10 V/m Recommended
separation distance
80MHz to 800MHz
800MHz to 2.7GHz