Instruction Manual

Guidance and manufacturer´s declaration – electromagnetic immunity
The Model RES5102 is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the Model RES5102
should assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
±2 kV, ±4 kV,
±8 kV,
±15 kV air
± 8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 kV air
Floors should be
wood, concrete
or ceramic tile.
If floors are
covered with
synthetic mate-
rial, the relative
humidity should
be at least 30%.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
100 kHz repeti-
tion frequency
± 1 kV for input/
output lines
N/A N/A
Surge
IEC 61000-4-5
± 0.5 kV, ± 1
kV differential
mode line-line
N/A N/A