Owner's Manual
Table Of Contents
- Preface
- Theory of Operation
- Specifications:
- Component Description
- Maintenance / Cleaning
- Returns
- Disposal Instructions
- Troubleshooting
- Sieve Bed Assembly Replacement (505849):
- Control Panel Replacement (507646):
- PC Board Replacement (506729/506729S O2 Sensor Models):
- Oxygen Sensor Replacement (508419) Models with O2 Sensor
- Internal Battery Replacement (507547):
- Compressor Plate Assy Replacement (506320):
- Checking the Oxygen Purity
17 | P a g e
Internal Battery Replacement (507547):
Tools and equipment required:
#1 Phillips screwdriver
Small diagonal wire cutters
Needle nose pliers
Cable Ties 4”x 3/32”
3/16” (4.7625 mm) ID silicone tubing x 6 inch (15 cm) long (minimum)
Disassembly
1. Follow steps 1 thru 11 of the “Disassembly” section of the
“Sieve Bed Replacement” in this manual.
2. Follow Steps 2 thru 7 of the “Disassembly” section of the
“Control Panel Replacement” in this manual.
3. Remove the 4 Screws securing Pump Compartment Cover
and set aside
4. Using diagonal cutters cut the Cable Tie that secures the
large rear Tube connecting the Inlet Muffler to the Pump and
remove the Tube from the Inlet Muffler.
5. Slide the Pump/Fan Assembly out of the POC, set aside.
6. Using diagonal cutters, cut the two (2) Cable Ties securing the
Battery to the Battery Shelf as shown and discard.
7. Remove the battery and dispose of in accordance to local recycling requirements.