Chapter 3 Installing the Electrical Supply 23
AC Source Site Requirements
Note As part of the installation, overcurrent devices meeting applicable national
and local electrical safety codes shall be provided.
The disconnect device for servicing is defined as any one of the following:
The appliance inlet on the rear of the system
The circuit breakers in the rack in which the system is mounted
The mains plug
It must be ensured that these remain accessible after installation.
Caution External filtering and/or surge suppression devices may be required on
the power feeds where branch circuit electromagnetic characteristics are unknown.
Chassis Enclosure Grounding
One Thomas & Betts two-hole lug (part number: 54204-UB) suitable for 8AWG
conductor or UL/CSA approved equivalent having 5/8-inch pitch. Torque value:
3.5Nm maximum. Two M5 studs and cupwasher nuts are supplied on the rear of
the chassis for connection.
A Thomas & Betts crimping tool (part number: TBM 5-S), or approved equivalent
is required to secure the lug on to the cable.
An earthing bus bar that is near the equipment and easily accessible.
The safety earth path is established by the connection of the grounding conductor
within the AC power cord to a reliably earthed socket outlet located near the
equipment.