User's Manual
130 SSCI - Suburban Surgical Company, Inc.
Chapter 5 - Repairs & Replacements
Filter, Filter/Fans
Compartment
P/N 853469
Procedure
1. Open the filter/fans compartment (refer to Page 58).
2. Bend the filter until the top or bottom clears the top or
bottom mounting bracket (Figure 153).
3. Remove the filter from the unit.
4. Fit the upper edge of the new filter to the upper filter
bracket.
5. Bend the filter slightly and fit the lower end into the lower
filter bracket.
6. Close and secure the filter/fans compartment panel.
Figure 153. Filter/Fans Compartment
Filter
Adjustment
Knob
Filter
Figure 154. Filter/Fans Compartment
Filter (removed from unit)