Product Manual

Note: the computer will recognize this unit when connecting in future uses.
You can obtain the latest rmware updates by contacting Subsurface Instruments
at info@ssilocators.com; create a folder in a convinient location, you will use this
folder to store the “*.HEX” (rmware update) le you receive from SSI.
APPENDIX
Figure 16-1
2. Click on the icon to run the
application.
3. The tiny bootloader screen will
display, (Figure 16-1).
4. The computer will automatically
recognize the Comm Port the AML+
is connected. The port # will
appear in the left panel of the screen.
5. Select your le source by clicking
on “Browse” button, select latest
rmware le and click “OK”
6. Go over the conguration settings in
all tabs, compare the values with the
values in page 17, make sure they
match.
7. On the left panel, click “Write Flash”.
This will start the transfer process
8. Upon completion, the screen will
display a string of values and
commands (Figure 17-4).
9. Your AML+ is now updated with the
latest rmware and needs to be
recalibrated.
Exit software.
1
2
1
2
16