Operating Instructions
Hermes READY TPS Consoles
REF 5100-201 and 5100-250
A Hermes-Interface cable links the two
systems. Linking the system does not
affect the normal function of the console.
At any time, console functions may be
selected from the control screen or by
using a control accessory such as a TPS
footswitch.
WARNINGS:
! Only the Hermes Operating Room
Control Center is approved for use
with the TPS console. Do not plug
any other device into the Hermes
console port. Use of other devices
may damage the console.
! It is important to read and understand
the Hermes Operating Room Control
Center instructions; the use of the
TPS console with the Hermes
Operating Room Control Center is
explained in those instructions.
! Stryker recommends attaching an
alternate control accessory
(footswitch) to the console in the event
the Hermes System fails during
surgery.
NOTE: The port labeled DATA PORT is
intended for the manufacturer's use only.
If you have questions about using the
Hermes System with the TPS console,
contact your Stryker Endoscopy
representative or call Stryker Endoscopy
Customer Service at 1-800-624-4422.
Outside the US, contact the nearest
Stryker subsidiary.
Connecting the Equipment
This is a system overview. For specific instructions on each TPS component, refer
to the information supplied with the component.
1. Place your console on a sturdy, flat surface near a hospital grade outlet.
2. Plug the console's power cord into the recessed power socket on the back of
the console.
3. Plug the other end of the power cord into a hospital-grade wall outlet.
4. Turn on the console. The on/off switch is located on the front of the console.
NOTE: As you set up the system, the console's screen will change to indicate the
various components as they are plugged in.
5. If using a Footswitch, plug the footswitch cable into the console port marked
FOOTSWITCH. Align orientation marks and gently push connectors together.
CAUTION: All TPS Cords have push/pull connectors. Do not thread or twist for
insertion or removal.
6. Plug the handpiece cord(s) into the console port(s) identified for the hand-
piece. Align connector orientation marks and gently push connectors together.
NOTE: Ports marked TPS1 and TPS2 are for Stryker TPS handpieces only.
The port marked ENDO is intended for the following list of Stryker Endoscopy
handpieces: SE5 Handpiece REF 272-704-100, SE5 Hand-controlled REF 272-
704; QuadraCut Shaver REF 275-701; QuadraCut Bone Plug REF 275-705; and
QuadraCut small Joint REF 275-601.
NOTE: If using a TPS Universal Handswitch, attach it to the handpiece before you
plug the cord into the handpiece.
7. Plug the other end of the handpiece cord(s) into the handpiece(s).
8. Attach cutting accessories to handpieces. Instructions supplied with each
handpiece or attachment provide details for cutting accessory assembly.
WARNING: Use only Stryker approved cutting accessories.
9. If using an Irrigation Console REF 5100-50 or 5100-250, assemble Irrigation
Pole REF 5100-50-28 to the console as shown. Hang irrigation bag from pole.
Install irrigation cassette into the pump. Attach irrigation clips to handpieces and
connect tubing.
Continue to the next step if using a Hermes READY™ TPS Console and Hermes
Operating Room Control Center.
CAUTION: Use only the Hermes-Interface cable provided by Stryker Endoscopy.
10. Connect the TPS console to the Hermes controller using a Hermes-Interface cable
into the port labeled Hermes on the back of the console. Plug the other end into
any available device port on the rear panel of the Hermes controller.
11. Power up both systems and test the devices to ensure they are performing properly
prior to surgery.
WARNINGS:
! Before using this system read and understand the information in this manual and the instructions supplied with each
TPS component and Stryker Endoscopy handpieces. Pay close attention to the User/Patient Safety Information.
! Familiarization with the Total Performance System prior to use is important. If you have any questions, contact your
Stryker Instruments representative or Stryker Customer Service at 1-800-253-3210.
! Prior to use, system components should be operated and inspected for any damage. DO NOT use if damage is
apparent.