User Manual
19
This screen appears only when using a TPS Footswitch
that is Revision 3 or newer. See Standard Features.
NOTE: The Footswitch revision is displayed on the SYSTEM
INFORMATION screen.
Access this screen by selecting the Footswitch icon on the
HANDPIECE CONTROL SCREEN or HANDPIECE ADJUST-
MENTS screen.
NOTE: These footswitch settings function only for the
handpiece for which they were selected. The handpiece is
identified at the top of the screen.
From this screen, the default functions of footswitch pads A
and B can be reprogrammed to accommodate the surgeon's
preferences. The left and right toggle icons correspond
respectively to footswitch pads A and B. Press each icon to
scroll through the following function options and stop on the
desired setting.
Off: Pedal is turned off.
Default: Footswitch defaults to the settings
selected on the handpiece screen.
Forward: Pedal setting default is overridden.
Pedal will provide handpiece rotation in the forward
(clockwise) direction.
Reverse: Pedal setting default is overridden.
Pedal will provide handpiece rotation in the reverse
(counterclockwise) direction.
Advanced: Use to access the Footswitch Button
Mapping screen. It will allow you to reprogram all
the footswitch pedals.
Return: Return to the previous screen.
Footswitch Adjustment Screen