Use Instructions
Table Of Contents
GGuuiiddaannccee aanndd MMaannuuffaaccttuurreerr’’ss DDeeccllaarraattiioonn -- EElleeccttrroommaaggnneettiicc IImmmmuunniittyy
Conducted RF
IEC 61000-4-6
3 Vrms150 kHz to 80
MHz outside ISM
bands 80% AM at 1
kHz
6 Vrms150 kHz to 80
MHz in ISM bands
80% AM at 1 kHz
3 Vrms150 kHz to 80
MHz outside ISM
bands 80% AM at 1
kHz
6 Vrms150 kHz to 80
MHz in ISM bands
80% AM at 1 kHz
WARNING: Portable
RF equipment
(including peripherals
such as antenna
cables and external
antennas) should be
used no closer than
30 cm (12 inches) to
any part of the
SurgiCount®+
System including
cables specified by
the manufacturer.
Otherwise,
degradation of the
performance of this
equipment could
result.
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey
1
should be
less than the
compliance level in
each frequency
range
2
.
Interference may
occur in the vicinity of
equipment marked
with the following
symbol:
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz to 2.7
GHz 80% AM at 1
kHz
3 V/m 80 MHz to 2.7
GHz 80% AM at 1
kHz
NNoottee
• At 80 MHz and 800 MHz the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios,amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the SurgiCount®+ System is
used exceeds the applicable RF compliance level above, the SurgiCount®+ System should be
observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the SurgiCount®+ System.
2
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
700001138880 Rev-AC 27 EN