Use Instructions

GGuuiiddaannccee aanndd MMaannuuffaaccttuurreerrss DDeeccllaarraattiioonn -- EElleeccttrroommaaggnneettiicc IImmmmuunniittyy
This product is intended for use in the electromagnetic environment specified below. The
customer or the user of this product should assure that it is used in such an environment.
IImmmmuunniittyy TTeesstt
IIEECC 6600660011 TTeesstt
LLeevveell
CCoommpplliiaannccee LLeevveell EElleeccttrroommaaggnneettiicc
EEnnvviirroonnmmeenntt --
GGuuiiddaannccee
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±2, ±4, ±8, ±15 kV
Air
±8 kV Contact
±2, ±4, ±8, ±15 kV
Air
±8 kV Contact
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
10%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV at 100 kHz
repetition frequency
for power supply
lines
±2 kV at 100 kHz
repetition frequency
for power supply
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±0.5, ±1 kV line(s) to
line(s)
±0.5, ±1, ±2 kV line
(s) to earth
±0.5, ±1 kV line(s) to
line(s)
±0.5, ±1, ±2 kV line
(s) to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage Dips, Short
Interruptions, and
Voltage Variations on
Power Supply Input
Lines
IEC 61000-4-11
0% U
T
(100% dip in
U
T
)
for 0.5 cycle at 0°,
45°, 90°, 135°,
180°, 225°, 270°, and
315°
0% U
T
(100% dip in
U
T
)
for 1 cycle at
70% U
T
(30% dip in
U
T
)
for 25 & 30 cycles at
0% U
T
(100% dip in
U
T
)
for 0.5 cycle at 0°,
45°, 90°, 135°,
180°, 225°, 270°, and
315°
0% U
T
(100% dip in
U
T
)
for 1 cycle at
70% U
T
(30% dip in
U
T
)
for 25 & 30 cycles at
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
30 A/m at 60 Hz 30 A/m at 60 Hz
Power frequency
magnetic fields
should be at levels
characteristics of a
typical location in a
typical commercial or
hospital environment.
EN 26 700001138880 Rev-AC