Use Instructions
Table Of Contents
EElleeccttrroommaaggnneettiicc CCoommppaattiibbiilliittyy
WWAARRNNIINNGG
• Avoid locating equipment adjacent to the SurgiCount®+ System. If such a configuration is
necessary, verify normal operation of both the SurgiCount®+ System and the adjacent
equipment during use.
• Always maintain a minimum separation distance of 6 inches [15 cm] from an in-use pacemaker
and the UHF RFID antenna of the SurgiCount®+ Reader while in use. Failure to comply may
cause interference with the pacemaker.
GGuuiiddaannccee aanndd MMaannuuffaaccttuurreerr’’ss DDeeccllaarraattiioonn -- EElleeccttrroommaaggnneettiicc EEmmiissssiioonnss
This equipment is intended for use in the electromagnetic environment specified below. The
customer or the user of this equipment should assure that it is used in such an environment.
EEmmiissssiioonnss TTeesstt CCoommpplliiaannccee EElleeccttrroommaaggnneettiicc EEnnvviirroonnmmeenntt -- GGuuiiddaannccee
FCC Part 15
WWAARRNNIINNGG -- Always maintain a minimum separation
distance of 9 inches [23 cm] from the head and torso of
a person and the UHF RFID antenna of the
SurgiCount®+ Reader while in use. Failure to comply
may cause FCC RF exposure limits to be exceeded.
The SurgiCount®+ System complies with FCC
regulations for conducted and radiated emissions under
FCC Part 15.
RF Emissions
CISPR 11
Group 2
WWAARRNNIINNGG -- The SurgiCount®+ System must emit
electromagnetic energy in order to perform its intended
function. Nearby electronic equipment may be affected.
RF Emissions
CISPR 11
Class A
WWAARRNNIINNGG -- This equipment is intended for use in a
professional healthcare facility environment. This
equipment may cause radio interference or may disrupt
the operation of other nearby equipment. It may be
necessary to take mitigation measures, such as
reorienting or relocating the SurgiCount®+ System or
shielding the location.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations/
Flicker
Emissions
IEC 61000-3-3
Complies
700001138880 Rev-AC 25 EN