Use Instructions
Table Of Contents
AAfftteerr UUssee
CClleeaanniinngg aanndd DDiissiinnffeeccttiioonn
WWAARRNNIINNGG
• Always clean and disinfect the equipment as indicated upon initial receipt and before each use.
Failure to comply may cause infection and result in patient or healthcare staff injury.
• Always consult the instructions for use that accompanies accessories for product specific
cleaning requirements.
CCAAUUTTIIOONN
• Do not immerse the equipment in liquid.
• Do not allow liquids or moisture to enter any electrical connection.
• Do not sterilize the equipment.
• Do not use solvents, lubricants, or other chemicals, including glutaraldehyde or similar
chemical cleaners, unless otherwise specified.
• Use of unapproved disinfectants may cause damage to equipment.
• Do not clean the reader when the battery cover is removed.
Failure to comply may cause damage to the equipment or degrade performance.
RReeccoommmmeennddeedd MMaatteerriiaallss
PPE as recommended by the disinfectant manufacturer
Soft, lint-free cloth
United States Environmental Protection Agency (US EPA) registered disinfectant with a claim
for activity against Hepatitis B. The following disinfectants have been validated for use on the
exterior surfaces of the equipment:
• Quaternary Ammonium Based – CaviCide (EPA #: 46781-6)
• Clorox Healthcare® Bleach Germicidal Wipes (EPA Reg. # 67619-12)
1. Wipe all external surfaces of the equipment with a soft, lint-free cloth moistened with a non-
abrasive, hospital disinfectant prepared according to the manufacturer’s instructions. Clean
surfaces until all visible soil is removed.
2. After removing all visible gross soil, use a clean cloth moistened with disinfectant and wipe all
surfaces. Make sure all surfaces remain visibly wet at room temperature for at least the
minimum time specified in the instructions supplied by the disinfectant manufacturer.
3. Remove any excess disinfectant solution using a soft, lint-free cloth moistened with water if
required by the instructions supplied by the disinfectant manufacturer.
CCAAUUTTIIOONN
• Do not use an aerosol spray directly on the reader barcode scanner window. Failure to
comply may result in damage to the device.
• Do not let liquid pool around the reader barcode scanner window. Failure to comply may
result in degraded performance.
4. Apply glass cleaner to a soft, lint-free cloth and clean the reader barcode scanner window.