User's Manual

EN-3
DRAFT
Please read this manual and follow its instructions carefullyThe words warning, caution, and note
carry special meanings and should be carefully reviewed:
Warning: Indicates risks to the safety of the patient or user� Failure to follow warnings may result
in injury to the patient or user
Caution: Indicates risks to the equipment� Failure to follow cautions may result in product
damage�
Note: Provides special information to clarify instructions or present additional useful information�
Warnings
Operator Prole
1� Federal (USA) law restricts this device to sale by or on the order of a physician�
2� The operator of the CrossFlow™ system should be a qualied health care professional having complete
knowledge of the use of this equipment and awareness of the risks associated with arthroscopic
procedures�
3� The operator of the system should be experienced in arthroscopic practices and techniques
4� The operator of the system should read this manual thoroughly and be familiar with its contents prior to
operating the equipment�
Prior to Surgery
1� Carefully unpack the CrossFlow Integrated Arthroscopy Pump and ensure that all components listed in
the “Package Contents section of this manual are accounted for and remain undamaged from shipment�
If damage to any component is detected, refer to the standard warranty
2� Install the system in an operating room that complies with all applicable IEC, CEC, and NEC requirements
for safety of electrical devices�
3� Install and use the system according to the information provided in the “Electromagnetic Compatibility”
section of this manual�
4� If the pump is installed with other equipment in a stacked conguration, observe the pump to verify
normal operation�
5� Portable and mobile RF communications equipment may interrupt system operation� When the
pump is in use, the conducted and radiated electrical elds may interfere with other electrical medical
equipment� If this occurs, power down all electrical equipment not in use, increase distance of other
electrical equipment, and/or connect the pump and other equipment into dierent outlets
6� Place the pump at the same height as the joint to ensure accurate pressure readings.
7� Ensure the proper connection of the primary power cord of the pump to a grounded receptacle with
the correct mains voltage� To prevent the risk of electric shock, do not use extension cords or portable
multiple socket outlets that are not a part of a certied hospital cart� The use of a portable multiple
socket outlet can lead to a reduced level of safety
8� Position any cables extending from the pump to avoid contact with the patient, electrodes, other cables,
and any electrical leads which provide paths for high frequency current�
9� Do not connect items which are not specied as part of the system� The use of accessories, transducers,
and cables other than those specied in this manual may result in increased emissions, decreased
immunity of the equipment, or unintended, unsafe operation of the system�
10� Examine all electrical connections to the pump before use� Improper connection may result in
malfunction or unintended surgical eects�
11� Do not touch or insert any objects, other than the cassettes, inside of the cassette holders as this may