Guidance and Manufacturer’s Declaration--Electromagnetic Immunity
The CROSSFLOW INTEGRATED ARTHROSCOPY PUMP system is intended for use in the electromagnetic environment specied below� The user of the CROSSFLOW INTEGRATED
ARTHROSCOPY PUMP system should ensure that it is used in such an environment�
Immunity Test
IEC 60601 Test
Level
Compliance Level Electromagnetic Environment--Guidance
Portable and mobile RF communications equipment should be used no closer to any part of the CROSSFLOW
INTEGRATED ARTHROSCOPY PUMP system, including its cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter
Conducted RF
IEC 61000-4-6
3Vrms
150kHz to 80MHz
3Vrms
Recommended Separation Distance:
d = 1�2√P
Radiated RF
IEC 61000-4-3
3V/m
80MHz to 2�5GHz
3V/m d = 1�2√P 80MHz to 800MHz
d = 2�3√P 800MHz to 2�5GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in meters (m)�
Field strengths from xed RF transmitters, as determined by an electromagnetic site survey (a), should be less
that the compliance level in each frequency range (b)�
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies�
NOTE 2: These guidelines may not apply in all situations� Electromagnetic propagation is aected by absorption and reection from structures, objects, and people
(a) Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and
TV broadcast cannot be predicted theoretically with accuracy� To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be
considered� If the measured eld strength in the location in which the CROSSFLOW INTEGRATED ARTHROSCOPY PUMP system is used exceeds the applicable RF compliance level
above, the CROSSFLOW INTEGRATED ARTHROSCOPY PUMP system should be observed to verify normal operation� If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the CROSSFLOW INTEGRATED ARTHROSCOPY PUMP unit�
(b) Over the frequency range 150kHz to 80MHz, eld strengths should be less than 3V/m�
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the CROSSFLOW INTEGRATED ARTHROSCOPY PUMP System
The CROSSFLOW INTEGRATED ARTHROSCOPY PUMP system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled� The user of
the CROSSFLOW INTEGRATED ARTHROSCOPY PUMP system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the CROSSFLOW INTEGRATED ARTHROSCOPY PUMP system as recommended below, according to the maximum output power of
the communications equipment�
Rated maximum output power
(W) of transmitter
Separation distance (m) according to frequency of transmitter
150kHz to 80MHz
d = 1�2√P
80kHz to 800MHz
d = 1�2√P
800kHz to 2�5GHz
d = 2�3√P
0�01 0�12 0�12 0�23
0�1 0�38 0�38 0�73
1 1�2 1�2 2�3
10 3�8 3�8 7�3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer
NOTE 1: At 80MHz and 800MHz, the separation distance for the higher frequency range applies�
NOTE 2: These guidelines may not apply in all situations� Electromagnetic propagation is aected by absorption and reection from structures, objects, and people
Regulatory Information
Federal Communications Commission (FCC)
FCC ID: SSH-XFLOW
Trade Name: CrossFlow Integrated Arthroscopy Pump
Type or Model: 0450000000
This device complies with Part 15 of the FCC rules� Operation is subject to the following two conditions:
this device may not cause harmful interference, and
this device must accept any interference received, including interference that may cause undesired
operation�
Note: FCC regulations provide that changes or modications not expressly approved by Stryker Endoscopy
could void your authority to operate this equipment