User's Manual

EN-32
DRAFT
Cleaning, Maintenance, and Disposal
Warning
Do not remove the cover of the console as this could cause electric shock and product damage.
To avoid electric shock and potentially fatal injury, unplug the pump from the electrical outlet
before cleaning
Caution
To prevent product damage:
when cleaning the pump, do not spray cleaning liquid directly onto the pump; spray on the cloth
before wiping the pump,
do not immerse the pump in any liquid,
do not use corrosive cleaning solutions to clean the pump,
do not sterilize the pump
Clean the Components
Pump
Should the pump need cleaning:
1� Spray cleaning liquid onto a dry, sterile cloth� Avoid excess liquid or drips�
2� Wipe the pump
3� Take extra care when cleaning the front LCD screen� Excess liquid or drips that enter the bottom of the screen
may result in product damage�
Footswitch
Warning
Clean the footswitch prior to rst use and after every subsequent use�
Consult the footswitch manual (P17862) for cleaning instructions�
Hand control
Warning
Clean and sterilize the hand control prior to rst use and after every subsequent use to minimize risk
of infection�
Consult the hand control manual for cleaning and reprocessing instructions�
Replace the Fuses
1� Disconnect the pump from the electrical output and remove the power cord from the rear of the pump
2� Remove the appropriate fuse holder; with a straight blade screwdriver, unlatch the fuse drawer, or turn the
global fuse holders in a counter clockwise direction until the spring pushes out�
3� Remove the fuse(s)�
4� Replace the fuse(s) with a fuse of the same value and rating as indicated on the rear of the pump�