User's Manual

EN-17
DRAFT
3� Select the scope/cannula combination
for option
;
to scroll up and down the
hardware combination list, press
and
4� Press the desired hardware combination to
highlight it�
In this scenario, “5�8mmx140mm cannula
with 4�0mm scope is highlighted� (Select
Low, Medium, or High Flow if the desired
hardware combination does not appear on
the list�)
5� If no other hardware combination options are
required, proceed to step 6�
6� If necessary, press
,
, and/or
to view the available scope/cannula
combinations for these options
Note: Hardware
and
are only
enabled if the hip joint is selected�
7� Repeat steps 3 and 4 to select the scope/
cannula combination for options
,
,
and/or
In this scenario, “4�0x120mm cannula with
2�7mm scope is highlighted for option
8� When the desired hardware combinations
are highlighted, press
to conrm the
selections and advance to the Run screen�
Swapping Between Hardware Combination Selections During the Case
The “Hot Swap function allows the user to switch the cannula through which the inow tubing is attached
without requiring recalibration� Depending on the surgical site, up to four cannulas can be utilized by this
function�
Warning
Use the scope and cannula as selected on the pump� An incorrect scope and cannula selection may
cause overpressure in the joint�