User's Manual

EN-13
DRAFT
1� Align the colored side of the cassette(s)
with the colored button(s) on the pump as
depicted in the diagram� The Inow Cassette
is blue and the Outow Cassette is red�
2� Insert the cassette(s)� Push with your thumb
until it clicks into place
When the Inow Cassette is inserted,
a green check mark will appear on the
cassette, and the screen will advance to
the User Preference File Selection screen�
When the Outow Cassette is inserted,
“Outow” will appear at the bottom of the
screen� A green check mark will appear
on the cassette if it is inserted before the
Inow Cassette (depicted in this scenario)�
3� Connect the tubing:
For the Inow and Outow Cassette
Tubing, refer to the instructions in the
CrossFlow Inow and Outow Cassette
Tubing manual�
For the Day-Use Inow Cassette
and Patient-Use Tubing, refer to the
instructions in the CrossFlow Day-Use
Inow Cassette and Patient-Use Tubing
manual�
Select the User Preference File
User preference les can be selected through the CrossFlow Pump interface� Select from “Standard Settings”
provided with the pump, or contact your Stryker representative to program your own, including settings for
pressure and ow rate and button assignments for the hand control and footswitch�