User's Manual
EN-57
DRAFT
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The Crossre2 System is intended for use in the electromagnetic environment specied below. The
customer or the user of Crossre2 System should ensure that it is used in such an environment
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic
Environment: Guidance
Electrostatic
Discharge (ESD) IEC
61000-4-2
±6kV contact
±8kV air
±2,4,6kV contact ±2,4,8kV
air
Floors should be wood,
concrete, or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst IEC
61000-4-4
±2kV for power supply
lines ±1kV for input/
output lines
±2kV for power supply
lines ±1kV for input/
output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV dierential mode
±2kV common mode
±0.5, 1kV dierential
mode ±1, 2kV common
mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut) for
5 cycles
70% Ut (30% dip in Ut) for
25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut) for
25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of Crossre2 System
requires continued
operation during power
mains interruptions,
it is recommended
that Crossre2 System
be powered from an
uninterruptible power
supply or a battery.
Power frequency
(50/60Hz) magnetic
eld IEC 61000-4-8
3 A/m N/A Power-frequency
magnetic elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.