User's Manual

EN-44
DRAFT
Troubleshooting
Problem Possible Solution
Console A hardware fault is
detected
Turn the power o and on again.
If the problem persists, contact a
Stryker representative or return
the console for repair.
The AC voltage is
incorrect
Turn the power off and on again.
If the problem persists, contact a
Stryker representative or return
the console for repair.
A software fault is
detected
Turn the power off and on again.
If the problem persists, contact a
Stryker representative or return
the console for repair.
The system does
not power on
Check the power cord to ensure it
is properly connected.
Check to ensure the cord is
connected to a grounded outlet.
The electrical
interference is
sporadic
Power down all electrical
equipment not in use.
Increase distance of other
electrical equipment.
Connect the unit and other
equipment into dierent outlets.
The generator
temperature is too
high
Ensure that there is proper airow
around the unit.
A power-on self
test error has
occurred
Turn the power o and on again.
If the problem persists, contact a
Stryker representative or return
the console for repair.