User Manual

26
TECHNICAL SPECIFICATIONS TABLES
The accuracy is specied for one year after calibration and at 18 to 28°C with a relative
humidity of 80%.
Overload Protection: 250V RMS. For 200mV range and 600V dc or RMS.ac for other ranges.
Overload Protection F 200mN250V fuse. (10 A range unfused)
Overload protection: 600V dc or RMS ac for all ranges.
Frequency range: 40Hz to 400Hz.
Response: Average responding, calibrated in RMS of a sine wave.
Range Resolution Accuracy
200mV 100μV ±0.5% of rdg ± 2 digits
2V 1mV ±0.5% of rdg ± 2 digits
20V 10mV ±0.5% of rdg ± 2 digits
200V 100mV ±0.5% of rdg ± 2 digits
600V 1V ±0.8% of rdg ± 2 digits
Range Resolution Accuracy
200μA 0.1μA ±1% of rdg ± 2 digits
2mA 1μA ±1% of rdg ± 2 digits
20mA 10μA ±1% of rdg ± 2 digits
200mA 100μA ±1.5% of rdg ± 2 digits
10A 10mA ±3% of rdg ± 2 digits
Range Resolution Accuracy
200V 100mV ±1 .2% of rdg ± 10 digits
600V 1V ±1 .2% of rdg ± 10 digits
DC VOLTAGE
DC CURRENT
AC VOLTAGE