User's Manual Part 2

STRATEGIC Technologies Inc. Active Mode Installation
9. Check Strap Tightness: Check the strap tightness on the side opposite the PID. There
should not be a space of more than ½ inch or 1.4 centimetres between the strap and the skin.
Installation Completion
10. Test the PID Signal: Watch the PRU+ in test mode, to ensure that the signal from the PID is
being received. The PID code number will be displayed and the PRU+ will beep to indicate it
has received the signal from the assigned PID. Note: during the first 5 seconds after the PRU+
receives the signal there will be a “+” immediately in front of the PID number. This will
disappear after 5 seconds. If the PID signal is not received within 1 minute, remove the PID
and check that the battery is properly installed.
11. Reset the PID: When a signal has been received by the PRU+, use the LOCK’s PID button to
reset the alarms on the PID. Wait until the next signal from the PID. If the alarms has been
properly reset, the PID number will be displayed, with no icons following it.
12. Complete the Installation: On the PRU+, press the “A” button until “B=Exit” is displayed.
Press button “B” to exit. Note: the telephone is now connected to the PRU+. Connect the
telephone line to the wall and to the “LINE” jack on the PRU+. The PRU+ will call the Central
Computer. When “Call Successful” is displayed, installation is complete.