User's Manual

3-56 KleeYa - User Manual - Version 2
BASIC FUNCTIONS
LOAD
STATUS INFO The Reagent screen shows the status of individual Reagent Cartridges and reagent
racks:
Table 3-5: Status info
PROCEDURE/
SHOW
INFORMATION
1. Tap on the Load button in the main menu bar to show the Load screen.
2. Tap on the Reagents button below the Load button.
LOADING Before a Reagent Cartridge is utilized, it must first be prepared. The mandatory
instructions on the packaging box of the Reagent Cartridges must be strictly
followed.
1. Remove the desired Reagent Cartridge from the package.
Keeping the Reagent Cartridge in an upright position at all times.
2. Prepare the Reagent Cartridge according to the instruction for use of the
relevant assay for details.
3. Load one or two Reagent Cartridges into the reagent rack.
Symbol/Time Description
Please wait while the RFID label is read and the particles are
shaken up.
Shaking magnetic beads.
Name Name of the Reagent Cartridge.
No cartridge Empty position in the reagent rack.
Number Number of remaining test count (shown below the Reagent
Cartridge).
Error:
Expiration date or the on-board stability is expired.
No valid calibration curve is available.
The remaining test count of the Reagent Cartridge are
zero.
See error message in the information box on the right side!