User's Manual
3-44 KleeYa - User Manual - Version 2
BASIC FUNCTIONS
LOAD
TUBE AND
SAMPLE RACK
SPECIFICATIONS
All used tubes (sample test tubes, calibrator tubes, or control tubes) must fulfill the
following specifications:
Table 3-1: Tube specifications
A special sample rack (A12, A13, ... , C10) must be used for each type of tube (1-8)
so that the instrument can correctly dispense liquid from the tubes.
Each (new) standard sample rack can be customized to a specific sample rack type
according to its intended use/used tubes,
see chapter 5.3.1 on page 5-11
.
LED STATUS The LEDs above sample loading bay lanes show the status of the sample rack in the
lane:
Table 3-2: LED status
Tube height Tube diameter
Sample rack
type
1 92 - 100 mm 12 mm A12
2 92 - 100 mm 13 mm A13
3 92 - 100 mm 15 mm A15
4 92 - 100 mm 16 mm A16
5 75 mm (+ spacer in sample rack) 12 mm B12
6 75 mm (+ spacer in sample rack) 13 mm B13
7 75 mm (+ spacer in sample rack) 15 mm B15
8 75 mm (+ spacer in sample rack) 16 mm B16
9 30 mm (in 75 mm transfer tube) 10 mm C10
LED Lane Notice
Off Lane is empty
Sample rack (samples, calibrators,
or controls) is finished or unused
Slow flashing Sample rack can be loaded in the
selected lane
On Sample rack (samples, calibrators,
or controls) will scheduled or is
processed
Do not unload the
sample rack!
Fast flashing Error See error message!